Replacement for Orion Chloride Ion Selective Electrode "9617"
Replacement for Thermo Scientific Orion, Acumet, Hach & many more, Series Chloride, "9617", ISE
$ Please call (1) 403 291 5570 for latest price or email
A Chloride Ion Selective Electrode sensor specifically designed to measure 1.8 ppm to 35,500 ppm (5x10-5 M to 1M) concentrations of Cl-.
Appliations include, but not limited to: River and Tap Water, Plant Tisue, Soils, Boiler Feed Water, Blood Electrolytes/Clinical Analysis, Sweat, Urine, Cement, Plating Baths, Food Samples.
Electrodes Direct Supply a wide variety of cost effective electrodes, sensors, assemblies and accessories for pH, Contacting Conductivity, Toroidal Conductivity, ORP, Dissolved Oxygen, Optical Dissolved Oxygen, Potentiostatic Chlorine, Karl Fischer, Ion-Specific and Custom Designs. Direct Replacements for Laboratory, Titration, Biotech, and Industrial
applications are available for all major brands.
Electrodes Direct # 137
Specifications
Model: CLO1502
Slope: 56 + / - 3 mV / decade
Reproducibility: + / - 2%
Interferences: S-2, I-1, CN-1, Br-1
Complexation: Heavy metals
Temperature Range: to 80° C (if calibrated at that temperature)
Pressure Range: 0 to 70 psi
Response Time: 95% response in 30 seconds
Storage: Long term, store dry Short term, store in dilute chloride standard
Concentration Range: 5 x 10-5 M to 1M 1.8 ppm to 35,500 ppm
pH Range: 2 to 12 pH
Resistance: < 1 megaohm
Care, Maintenance & Cleaning: Polish with polishing strips Treat with a dilute ISA
Temperature Compensation Not recommended
Isopotential Point: 50,000 ppm Chloride
On-Line Capability: Yes
Applications: River and Tap Water, Plant Tissue, Soils, Boiler Feed Water, Blood Electrolytes/Clinical Analysis, Sweat, Urine, Blood, Cement, Plating Baths, Food Samples
Replaces, Thermo Scientific Orion, Acumet, Hach & many more, Chloride, "9617"